Skip to main content

Table 4 Precision (Intermediate assay and Intra-day) validation for dieckol in phlorotannin preparations

From: Method development and validation for dieckol in the standardization of phlorotannin preparations

 

Concentration (μg/mL)

Instruments

Dieckol content (mg/g)

Average content (mg/g)

SD (mg/g)

RSD (%)

Itermediate assay

94.56231

Agilent 1260

73.930

74.453

0.920

1.236

 

97.25632

Agilent 1260

75.057

 

93.42563

Agilent 1260

73.044

 

98.62358

Agilent 1200

75.384

 

Phlorotannin sample (mg)

Concentration (μg/mL)

Dieckol content (mg/g)

Average content (mg/g)

SD (mg/g)

RSD (%)

Intra-day: Agilent 1260 Infinitya

56.5

96.10084

74.1592

73.9297

0.761

1.030

 

59.4

101.08464

74.1968

 

58.3

97.25135

72.7300

 

56.6

95.55534

73.6080

 

57.9

99.64617

75.0358

 

52.7

89.26187

73.8485

  1. SD standard deviation, RSD relative standard deviation
  2. aValues were calculated using intra-day analyses (n = 6)